Model 6600-NG: RVSM Automated Pitot Static Tester (For SmartProbe™ Aircrafts)
The Model 6600-NG Automated Tester is designed to test aircraft equipped with SmartProbes™. The tester has one Static output (Ps1), one Pitot output (Pt) and a third output (Ps2) which can be used to simulate either AOA (angle of attack) or AOS (angle of sideslip). Using the small and light-weight Remote, an operator can test the entire pitot and static system from the cockpit. The Model 6600-NG comes with a removable 8-hour battery (optional), wide operating temperature (optional), and “profiles” to accelerate testing. Calibration is required only once a year. The 6600-NG is also WiFi-enabled, which enables wireless remote control using an iPad or other tablets.
The Model 6600-NG is approved for Embraer EJETS, Bombardier C-Series, Gulfstream, Honda Jet and most SmartProbe™ equipped aircraft.
- Specifications
- Features
- Documents
- Ordering Information
Description
Ps1 Output
Pressure function
Range:
0.1 to 32 or 42 inHg
Resolution:
0.001 inHg
Accuracy:
±0.002 inHg
Units:
inHG, mbar, psi, mmHG, kpa
Altitude function
Range:
-4000 ft. to 60,000 ft.
Resolution:
1 foot
Accuracy:
± 2 ft. @ 0 ft.
± 6 ft. @ 35,000 ft.
± 12 ft. @ 50,000 ft.
Climb function
Range:
0 ft/min to ±25,000 ft/min
Resolution:
1 ft/min
Accuracy:
±1% of rate of climb
Units:
Feet, meters, inHg, mbar, psi, kpa, mmHg
Ps2 Output
Pressure function:
Same as for Ps1
Altitude function:
Same as for Ps1
Climb function:
Same as for Ps1
Airspeed Function
Range:
0 to 500 knots
Resolution:
0.1 knots
Accuracy:
0.3 knots @ 50 knots
0.16 knots @100 knots
0.05 knots@ 500 knots
Displayed Units
(Diff. or Abs.) inHg, mbar, psi, kpa, mmHg, feet, meters, knots, kmph
Pitot Output
Pressure Function
Range:
0 to 30 inHg(diff.)
Resolution:
0.0001 inHg
Accuracy:
±0.003 inHg
Airspeed function
Range:
0 to 690 knots/min
Resolution:
0.1 knots
Accuracy:
0.3 knots @ 50 knots
0.2 knots @100 knots
0.05 knots@ 690 knots
Mach Function
Range:
0.0 to 3.0 Mach
Resolution:
0.001 Mach
Accuracy:
±0.001 above 0.2 Mach
Displayed Units
knots, Mach, kmph, mph, EPR (Diff. or Abs.) inHg, mbar, psi, kpa, mmHg
Power Requirement
90-260 VAC, 47-440 Hz, 100 W (200 W with heaters)
Optional Removable Battery: 6-8 hour operation
External Interfaces
Wi-Fi, RS232 standard / Encoder optional
Dimensions & weights
Main unit:
22” x 14” x 9” / 38 lbs
Remote unit:
6” x 6” x 1” / 1.8 lb.
Environmental Specs
Operating temp.
0°C to 50°C, –40°C to 50°C (optional)
Storage temp.:
–40°C to 75°C
Humidity:
5 to 100%
Features
Three Outputs In addition to the Primary Static and Pitot outputs, a third output is provided. This output can be used as a secondary static, a differential static or secondary pitot. This output allows the user to create a second altitude (in feet or mtrs), a second absolute pressure (in inHg or mbar), a differential pressure for simulating AOA (in Diff. inHg or Diff mbar) or a differential pressure for generating a second airspeed (in knots or kmph) to simulate AOS. When the third output is not used, the tester operates exactly like our 2-output tester (Model 6300) |
WiFi-enabled |
RVSM compliant The high accuracy of the tester and NIST-traceable calibration make the tester fully RVSM-compliant and suitable for use on all RVSM-compliant aircraft. |
Calibration period: 1 year |
Automated operation Simply enter the target set-points for the three outputs and the unit automatically achieves the targets, ramping at the programmed rates. Control at the set-point is stable within ± 1 foot, ± 0.1 knot or ± 0.001 inHg |
Programmable limits User-programmable limits for altitude, airspeed, differential pressure, Mach and VSI provide protection to the aircraft. |
“Jog” and “Hold” features Set-points can be “jogged” up or down in small increments. Ramping to a set-point can be put on “hold” and controlled at the current measured value. |
“Ground” feature Brings Pitot and Static outputs to ambient “ground” pressure after completing a test. |
“Profiles” feature A profile of set-points can be downloaded from a PC, allowing the user to step through the set-points using a single key. Each profile can have up to 50 set-points and 12 profiles can be stored in the unit. |
Leak checks Leak-checks can be performed independently on any of the three outputs. The display of “instant” leak rates provides immediate feedback to the user. |
Pressure & Vacuum source Built-in high-capacity dual vacuum pumps and pressure pump provide the pressure and vacuum source for the unit. |
Remote unit The Remote unit is the operator interface for the tester. All valid parameters of all three outputs are clearly displayed simultaneously on a single screen on the remote. |
Manual Vent The tester includes precision metering valves to enable manual venting of all three outputs, in the event of power-loss |
Applications
Ideal tester for use with aircraft equipped with SmartProbe™ that have the ability to sense AOA and/ or AOS. |
Leak-checks and Air-data-system accuracy-checks on all RVSM aircraft including aircraft with SmartProbe™. The tester can be used either in 3-output mode or 2-output mode. In 2-output mode this tester operates exactly like our Model 6300. |
System
Accessories