Model 6500/6500-HA: Automated Air Data Test Set
NSN: 4920 01 375 7472 (Model 6500)
The 6500 & 6500-HA (High-Accuracy) Air Data Test Sets require only a pressure source and a vacuum source and are extremely simple to use. They can be used to test and calibrate altimeters, airspeed / Mach indicators, Climb indicators, Flight data recorders, Air data computers, and EPR indicators. The high accuracy of these units meets all RVSM requirements. Also, they need to be calibrated only once a year. The use of “Profiles” makes it possible for the operator to run through a test using only a single key on the front panel keypad of these units.
- Specifications
- Features
- Documents
- Ordering Information
Description
Static Output
6500
6500-HA
Pressure function
Range:
0.1 to 42 inHg
0.1 to 38 inHg
Resolution:
0.001 inHg
0.0001 inHg
Accuracy:
±0.002 inHg
±0.001 inHg
Units:
inHg, mbar
inHg, mbar
Altitude function
Range:
-4000 ft. to 100,000 ft.
-4000 ft. to 100,000 ft.
Resolution:
1 foot
1 foot
Accuracy:
± 2 ft. @ 0 ft.
± 6 ft. @ 35,000 ft.
± 12 ft. @ 50,000 ft.
± 50 ft. @ 80,000 ft.
± 1 ft. @ 0 ft.
± 3 ft. @ 35,000 ft.
± 6 ft. @ 50,000 ft.
± 25 ft. @ 80,000 ft.
Units:
Feet, Meters
Feet, Meters
Climb function
Range:
0 ft/min to ±50,000 ft/min
0 ft/min to ±50,000 ft/min
Resolution:
1 ft/min
1 ft/min
Accuracy:
±1% of rate of climb
±1% of rate of climb
Units:
Feet/min, Meters/sec
Feet/min, Meters/sec
Pitot Output
Pressure Function
Range:
0.1 to 110 inHg
0.1 to 100 inHg
Resolution:
0.001 inHg
0.0001 inHg
Accuracy:
±0.003 inHg
±0.002 inHg
Pt Units:
inHg, mbar, EPR
inHg, mbar, EPR
Qc Units:
inHg, mbar
inHg, mbar
Airspeed Function
Range:
0 to 1100 knots
0 to 1100 knots
Resolution:
0.1 knots
0.1 knots
Accuracy:
± 1.5 knots @ 20 knots
± 0.5 knots @ 50 knots
0.25 knots above 100 knots
0.1 knots above 300 knots
± 1.0 knots @ 20 knots
± 0.2 knots @ 100 knots
± 0.03 knots @ 500 knots
± 0.01 knots @ 1000 knots
Units:
knots, Mach, km/hr, mph
knots, Mach, km/hr, mph
Rate of change:
0 to 500 knots/min
0 to 500 knots/min
Mach Function
Range:
0.0 to 5.0 Mach
0.0 to 10.0 Mach
Resolution:
0.001 Mach
0.001 Mach
Accuracy:
±0.001 above 0.2 Mach
±0.001 above 0.2 Mach
Power Requirement
90-260 VAC, 47-440 Hz ; 200 VA
External Interfaces
RS232 and IEEE-488; Remote unit optional
Dimensions & weights
19-inch rack-mount kit optional: 6.5” (h) x 15.5” (w) x 17.0” (d) / 20 lbs
Environmental Specs
Operating temp.
0 to 50°C
0 to 50°C
Storage temp.:
-25 to 75°C
-25 to 75°C
Humidity:
5 to 95% non-condensing
5 to 95% non-condensing
Features
RVSM compliant 6500: The high accuracy of the tester and NIST-traceable calibration make the tester fully RVSM-compliant. 6500-HA: The very high accuracy of the tester and NIST-traceable calibration make the tester fully RVSM-compliant. |
Calibration period 6500/6500-HA: 1 year |
Automated operation 6500: Simply enter the target set-points for altitude and airspeed and the unit automatically achieves the targets, ramping at the programmed rates. Control at the set-point is stable within ± 1 foot, ± 0.1 knot or ±0.001 inHg. 6500-HA: Simply enter the target set-points for altitude and airspeed and the unit automatically achieves the targets, ramping at the programmed rates. Control at the set-point is stable within ± 1 foot, ± 0.1 knot or ± 0.0002 inHg |
Programmable limits 6500/6500-HA: User-programmable limits for altitude, airspeed, Mach and VSI provide protection to the aircraft. |
“Jog” and “Hold” features 6500/6500-HA: Set-points can be “jogged” up or down in small increments. Ramping to a set-point can be put on “hold” and controlled at the current measured value. |
“Ground” feature 6500-HA: Brings Pitot and Static outputs to ambient “ground” pressure after completing a test. |
“Profiles” feature 6500/6500-HA: A profile of set-points can be downloaded from a PC, allowing the user to step through the set-points using a single key. Each profile can have up to 50 set-points and 20 profiles can be stored in the unit. |
Leak checks 6500/6500-HA: Pitot and Static leak-checks can be performed either independently or simultaneously. Three leak-timers can be programmed to perform timed leaks between 1 to 20 minutes. The display of “instant” leak rates provides immediate feedback to the user. |
Pressure & Vacuum requirement 6500/6500-HA: External pressure (Dry Air only) and vacuum supplies are required. The PVS is ideally suited for providing these supplies. |
Local display and keypad 6500/6500-HA: The local display and keypad are the operator interface for the tester. All valid parameters, including altitude, climb and airspeed, are clearly displayed simultaneously on a single screen on the local display. |
Applications
6500/6500-HA: Ideal tester for use in an instrument-shop for testing and calibrating ADC’s, altimeters, airspeed indicators, VSI’s, cabin pressure controllers, FDR’s and other air-data instruments. |
6500/6500-HA: Can be rack-mounted in ATE systems |
6500-HA: The very high accuracy of this tester makes it suitable for use as a convenient Transfer Standard. |
System
Accessories