Specifications

Static Output

Pressure function

Range:
Resolution:
Accuracy:
Units:
1 to 32 inHg
0.001 inHg
±0.005 inHg
inHG, mbar, psi, mmHG, kpa

Pressure Rate

Range:
Resolution:
Accuracy:
Units:
0 to 30 inHg/min
0.001 inHg/min
±1% of rate
pressure-units/min

 

Altitude function

Range:
Resolution:
Accuracy:
-2000 ft. to 50,000 ft.
1 foot
± 5 ft. @ 0 ft.
± 6.5 ft. @ 10,000 ft.
± 10 ft. @ 20,000 ft.
± 15 ft. @ 35,000 ft.
± 30 ft. @ 50,000 ft.
Units: Feet, Meters

Altitude Rate function

Range:
Resolution:
Accuracy:
Units:
0 ft/min to ±10,000 ft/min
1 ft/min
±1% of rate of climb
Feet/min, Meters/sec

 

Pitot Output

Pressure function

Range:
Resolution:
Accuracy:
Pt units:
Qc units:
0 to 15 inHg (differential)
0.001 inHg
±0.003 inHg
inHg, mbar, psi, mmHg, kpa, EPR
inHg, mbar, psi, mmHg, kpa

Pressure Rate

Range:
Resolution:
Accuracy:
Units:
0 to 30 inHg/min
0.001 inHg/min
±1% of rate
pressure-units/min
(EPR in inHg/min)

Airspeed function

Range:
Resolution:
Accuracy:
0 to 500 knots
0.1 knots
± 1.0 knots @ 20 knots
± 0.3 knots @ 50 knots
± 0.2 knots @ 100 knots
± 0.1 knots @ 200 knots
± 0.04 knots @ 500 knots
Units: knots, Mach, km/hr, mph

Airspeed Rate

Range:
Resolution:
Accuracy:
Units:
0 to 500 knots/min
0.1 knots/min (0.01 km/hr/sec)
±2% of rate
knots/min, km/hr/sec, mph/min

Mach function

Range:
Resolution:
Accuracy:
0.0 to 1.0 Mach
0.001 Mach
±0.001 above 0.2 Mach

Power requirement

90-260 VAC, 47-440 Hz, 100 W (200 W with heaters)
Optional Internal Battery: 8 hours of operation

External Interfaces

WiFi, RS232 Serial ; Encoder optional

Dimensions & weights

Main unit:
Remote unit:
22” x 14” x 9” / 30 lbs
6” x 6” x 1” / 1.8 lb.

 

Environmental specs

Operating temp.
Storage temp.
Humidity:
0°C to 50°C, –40°C to 50°C (optional)
–40°C to 75°C
5 to 100%

Features

WiFi-enabled

Automated operation (at an Affordable price) Simply enter the target set-points for altitude and airspeed and the unit automatically achieves the targets, ramping at the programmed rates. Control at the set-point is stable within ± 1 foot and ± 0.1 knot. The tester’s low price makes it very affordable

Calibration period: 1 year

Programmable limits User-programmable limits for altitude, airspeed, Mach, VSI and airspeed-rate provide protection to the aircraft.

Negative-airspeed protection An internal differential-pressure relief valve provides protection against negative airspeed.

Internal Battery (Optional) The internal battery allows for 8 hours of continuous operation. Re-charge time is only 8 hours.

Low-temperature operation (optional) With optional internal heaters, the unit can be operated down to -40°C

“Jog” and “Hold” features Set-points can be “jogged” up or down in small increments. Ramping to a set-point can be put on “hold” and controlled at the current measured value.

“Ground” feature Brings Pitot and Static outputs to ambient “ground” pressure after completing a test. Ambient pressure is constantly monitored.

“Profiles” featureA profile of set-points can be downloaded from a PC, allowing the user to step through the set-points using a single key. Each profile can have up to 50 set-points and 15 profiles can be stored in the unit.

Leak checks Pitot and Static leak-checks can be performed either independently or simultaneously. Three leak-timers can be programmed to perform timed leaks between 1 to 20 minutes. The display of “instant” leak rates provides immediate feedback to the user.

Pressure & Vacuum source Built-in diaphragm pumps provide the pressure and vacuum source for the unit.

Remote unitThe Remote unit is the operator interface for the tester. It features a sunlight-readable 5.8” color display, a back-lit keypad and a touch-screen (which can be disabled). All operations are performed from a single screen which displays all pertinent parameters.

Manual Vent Valves On loss of power, the unit maintains pressure (except for leaks). The unit can be manually vented to “Ground” at a controlled rate using precision metering valves.

Applications

  • Leak-checks and accuracy-checks on all non-RVSM aircraft, helicopters or UAVs.
  • Leak-checks and limited accuracy-checks on RVSM aircraft
  • Ideal for line-maintenance by commercial airlines